Specifications
Standard Lead Length: 1 m (40") Operating Temperature Range:
TFE Leads: -200 to 260°C
(-320 to 500°F)
Polyimide Leads: -200 to 350°C
(-320 to 660°F)
Fiberglass Leads: -75 to 510°C
(-100 to 950°F)
Moisture Resistant: -50 to 200°C
(-60 to 390°F)
Sensing Element: International grade thin film platinum, a = 0.00385 Ω/Ω/°C per IEC 60751
Ice Point Resistance: 100.00±0.06Ω or 1000.0±0.6Ω at 0°C
Time Constant (Water at 3'/Second):
0.086 Diameter = 0.5 seconds
0.098 Diameter = 0.6 seconds
0.125 Diameter = 0.8 seconds
0.187 Diameter = 1.5 seconds
Self-Heating: Greater than 15 mW/°C Interchangeability/Accuracy:
Class A (±0.15+.002*|t|°C [where
|t|= absolute value of temperature in °C]) between -50 and 300°C, Class B above and below these temperatures Long Term Stability: Better than 0.05°C (0.02% of resistance) per
5 years, -50 to 300°C (-58 to 572°F), 0.25°C per year to 540°C (1004°F)